FDA Adverse Event Injury Summary report: N

GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN

MDR report key: 2805417 · Received September 17, 2010

Report

Report Number
3003900188-2010-00035
Event Type
Injury
Date Received
September 17, 2010
Date of Event
May 28, 2010
Manufacturer
LINDE GAS NORTH AMERICA LLC
Product Code
ECX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

THE PRESENT CASE IS ONE OUT OF 9 INCIDENT REPORTS IN CONNECTION WITH 7 LIV IGNITIONS. EVAL SUMMARY: ROOT CAUSE ANALYSIS SUMMARY - THE CAUSE OF IGNITION WITH HIGHEST PROBABILITY IS IGNITION OF HYDROCARBONS, LUBRICANT AND/OR DEBRIS BY ADIABATIC COMPRESSION. POTENTIAL SOURCES OF HYDROCARBONS AND LUBRICANTS; LUBRICANT (FOMBLIN) MIGRATION (LUBRICANT FROM GCE MFG OF VALVE), LEACHING OF CONTAMINANTS (PHTHALATES) FROM O-RINGS IN VALVES (LUBRICANT IN SOLVENT) AND HYDROCARBON OIL, LEACHING OF CONTAMINANTS (PHTHALATES) FROM FILL CONNECTOR AT FILLING PLANT AT (B)(4). POTENTIAL SOURCES OF DEBRIS - DEBRIS FROM MFG PROCESS OF VALVE AT GCE, DEBRIS FROM FILLING PLANT AT LINDE, DEBRIS FROM WEAR, VALVES IN USE. THE DESIGN (FILTER AND LOCATION OF COMPONENTS IN THE HIGH PRESSURE SIDE OF THE VALVE) OF THE GCE LIV VALVE 5 MICRON FILTER IS LIKELY TO CAUSE MORE SENSITIVITY TO ADIABATIC COMPRESSION HEATING THAN THE GCE OXYGEN DESIGN.

Description of Event or Problem · 1

ON (B)(6) 2010, A MALE NURSE ((B)(6) YEARS OF AGE) DELIVERED OXYGEN TO A HOMECARE PT WITH A CHRONIC ASTHMA ATTACK, AND WHEN THE CANNULA WAS PLACED, THERE WAS A FLASH THAT BROKE THE TOMAGOMA OF THE CYLINDER. THE NURSE GOT A FIRST DEGREE BURN IN SIX FINGERS AND WAS OUT OF WORK FOR 10 DAYS. THE NURSE RECOVERED FROM THE BURNS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN NONE ECX LINDE GAS NORTH AMERICA LLC 0710160 4768666

Patients

Seq Age Sex Outcome Treatment
1