FDA Adverse Event Injury Summary report: N

GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN

MDR report key: 2805401 · Received September 17, 2010

Report

Report Number
3003900188-2010-00039
Event Type
Injury
Date Received
September 17, 2010
Date of Event
November 23, 2008
Manufacturer
LINDE GAS NORTH AMERICA LLC
Product Code
ECX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRESENT CASE IS ONE OUT OF 9 INCIDENT REPORTS IN CONNECTION WITH 7 LIV IGNITIONS. EVAL SUMMARY: ROOT CAUSE ANALYSIS SUMMARY - THE CAUSE OF IGNITION WITH HIGHEST PROBABILITY IS IGNITION OF HYDROCARBONS, LUBRICANT, AND/OR DEBRIS BY ADIABATIC COMPRESSION. POTENTIAL SOURCES OF HYDROCARBONS AND LUBRICANTS; LUBRICANT (FOMBLIN) MIGRATION (LUBRICANT FROM GCE MFG OF VALVE), LEACHING OF CONTAMINANTS (PHTHALATES) FROM O-RINGS IN VALVES (LUBRICANT IN SOLVENT) AND HYDROCARBON OIL, LEACHING OF CONTAMINANTS (PHTHALATES) FROM FILL CONNECTOR AT FILLING PLANT AT LINDE. POTENTIAL SOURCES OF DEBRIS - DEBRIS FROM MFG PROCESS OF VALVE AT GCE, DEBRIS FROM FILLING PLANT AT LINDE, DEBRIS FROM WEAR, VALVES IN USE. THE DESIGN (FILTER AND LOCATION OF COMPONENTS IN THE HIGH PRESSURE SIDE OF THE VALVE) OF THE GCE LIV VALVE 5 MICRON FILTER IS LIKELY TO CAUSE MORE SENSITIVITY TO ADIABATIC COMPRESSION HEATING THAN THE GCE OXYGEN DESIGN.

Description of Event or Problem · 1

ON (B)(6) 2008, LINDE WAS NOTIFIED OF AN INCIDENT IN CONNECTION WITH THE USE OF A LINDE INTEGRATED VALVE. ON (B)(6) 2008, ON ARRIVAL OF A PT AT (B)(6), THE ASSISTANT ANAESTHESIST WAS PREPARING TO GIVE THE PT MEDICINAL OXYGEN. THE LIV 2 L CYLINDER WAS MOUNTED ON THE HEAD OUTSIDE THE BED (VERTICAL POSITION). THE ASSISTANT ANAESTHESIST OPENED FIRST THE FLOW-REGULATOR TO 15 L/MIN AND OPENED THEN THE MAIN VALVE. WHILE CONNECTING THE HOSE/TUBE TO THE NIPPLE, AN AUDIBLE HISSING SOUND WAS HEARD INSIDE THE CAP, FOLLOWED BY A FLAME. ALL PERSONNEL PRESENT FLED FROM THE AREA AND THE CYLINDER FELL ON THE GROUND. SHORTLY AFTER THAT, A FLAME CAME OUT OF THE VALVE, WHICH EXTINGUISHED AFTER A SHORT TIME. THE FIRE-ALARM WAS ACTIVATED AND THE FIRE BRIGADE ARRIVED 8 MINUTES LATER. THERE WAS SOOT AT THE BEGINNING OF THE HOSE, THE MASK WAS CLEAN. DURING THE INCIDENT, THE PT WAS UNCONSCIOUS. THE ASSISTANT ANAESTHESIST SUFFERED OF SECOND DEGREE BURNS (RECUPERATABLE) ON HIS HANDS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN NONE ECX LINDE GAS NORTH AMERICA LLC 0710204 2551101

Patients

Seq Age Sex Outcome Treatment
1