FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 600 SYNCHRON SYSTEM

MDR report key: 2805245 · Received October 25, 2012

Report

Report Number
2050012-2012-01761
Event Type
Malfunction
Date Received
October 25, 2012
Date of Event
September 29, 2012
Report Date
September 30, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
JGS
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Description of Event or Problem · 1

CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT THE UNICEL DXC 600 SYNCHRON SYSTEM GENERATED ERRONEOUS SODIUM AND CHLORIDE RESULTS. CUSTOMER REPORTED THAT THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) REPLACED THE CARBON BRIDGE AND THE MODULAR CHEMISTRIES SAMPLE PROBE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXC 600 SYNCHRON SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JGS BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1