FDA Adverse Event
Malfunction
Summary report: N
UNICEL DXC 600 SYNCHRON SYSTEM
MDR report key: 2805245
·
Received October 25, 2012
Report
- Report Number
- 2050012-2012-01761
- Event Type
- Malfunction
- Date Received
- October 25, 2012
- Date of Event
- September 29, 2012
- Report Date
- September 30, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Description of Event or Problem · 1
CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT THE UNICEL DXC 600 SYNCHRON SYSTEM GENERATED ERRONEOUS SODIUM AND CHLORIDE RESULTS. CUSTOMER REPORTED THAT THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) REPLACED THE CARBON BRIDGE AND THE MODULAR CHEMISTRIES SAMPLE PROBE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 600 SYNCHRON SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JGS | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |