FDA Adverse Event Summary report: N

STERIS SYSTEM 1E

MDR report key: 2803291 · Received October 24, 2012

Report

Report Number
3000251274-2012-00236
Date Received
October 24, 2012
Date of Event
September 30, 2012
Report Date
October 24, 2012
Manufacturer
STERIS CORPORATION - HOPKINS
Product Code
MED
PMA / PMN Number
K090036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

STERIS IDENTIFIED THROUGH A PRODUCTION/POST PRODUCTION RISK MANAGEMENT REVIEW THAT PROPERTY DAMAGE CAN OCCUR IF A SYSTEM 1E HOSE DISCONNECTS, RESULTING IN WATER LEAKAGE WHEN THE UNIT IS LEFT ON AND UNATTENDED AT NIGHT AND/OR WEEKENDS. STERIS HAS RECEIVED NO REPORTS OF INJURIES ASSOCIATED WITH THE DISCONNECTION OF A SYSTEM 1E WATER HOSE. THE STERIS SERVICE TECHNICIAN HAS INSTALLED NEW HOSES AND CONNECTIONS ON THIS SYSTEM 1E PROCESSOR ((B)(4)). THE TECHNICIAN TESTED THE UNIT, INCLUDING RUNNING A DIAGNOSTIC AND PROCESSING CYCLE AND CONFIRMED THE UNIT WAS OPERATIONAL.

Description of Event or Problem · 1

THE USER FACILITY REPORTED A HOSE SEPARATED SPRAYING APPROXIMATELY 10 GALLONS OF WATER ONTO THE FLOOR OF THE ROOM WHERE THE SYSTEM 1E IS LOCATED AND CREATED A PUDDLE IN AN ADJACENT HALLWAY. THE WATER SUPPLY WAS SHUT OFF AND CLEANED UP BY FACILITIES MAINTENANCE. NO REPORT OF INJURY, PROCEDURAL DELAY OR PROPERTY DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERIS SYSTEM 1E LIQUID CHEMICAL STERILANT PROCESSING SYSTEM MED STERIS CORPORATION - HOPKINS

Patients

Seq Age Sex Outcome Treatment
1