FDA Adverse Event Malfunction Summary report: N

LARGE SUTURECUT NEEDLEDRIVE

MDR report key: 2803185 · Received October 24, 2012

Report

Report Number
2955842-2012-00652
Event Type
Malfunction
Date Received
October 24, 2012
Date of Event
September 5, 2012
Report Date
September 24, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING OBSERVED ONE GRIP CLOSE CABLE BEING BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPINS FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT STICKS OUT AT THE WRIST. OTHER CABLES AT WRIST ARE NOT DAMAGED. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE, THE WIRES OF THE LARGE SUTURECUT NEEDLEDRIVER INSTRUMENT BROKE WHILE SUTURING. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LARGE SUTURECUT NEEDLEDRIVE ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420296-02 M10120629 400

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES