FDA Adverse Event Summary report: N

MICRO DRILL MEDIUM STRAIGHT ATTACHMENT

MDR report key: 2803151 · Received October 17, 2012

Report

Report Number
1811755-2012-03827
Date Received
October 17, 2012
Date of Event
September 18, 2012
Report Date
October 4, 2012
Manufacturer
STRYKER INSTURMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CORROSION DAMAGE WAS NOTED WITHIN THE DEVICE.

Description of Event or Problem · 1

A MICRO DRILL MEDIUM STRAIGHT ATTACHMENT WAS SENT FOR EVALUATION BECAUSE A MICROAIRE BUR WAS STUCK IN IT. HOWEVER, DURING QUALITY TESTING, SEVERE WOBBLING WAS NOTED. NO ADVERSE CONSEQUENCE WAS ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO DRILL MEDIUM STRAIGHT ATTACHMENT ERL STRYKER INSTURMENTS KALAMAZOO 12101

Patients

Seq Age Sex Outcome Treatment
1 UNK