PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE
Report
- Report Number
- 1419937-2012-00647
- Event Type
- Malfunction
- Date Received
- October 19, 2012
- Date of Event
- September 18, 2012
- Report Date
- September 18, 2012
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. IN THE FOLLOW UP WITH THE CUSTOMER, SHE INDICATED THAT AFTER SHE CALLED IN WITH THE LOW SUCTION ISSUE RIGHT BEFORE SHE WAS GOING TO SEND HER TRANSFORMER BACK, SHE NOTICED THE CRACK IN THE TRANSFORMER HOUSING AND IT SPLIT APART. THE CUSTOMER ALSO INDICATED THAT SHE DID NOT WITNESS ANY SMOKE, FIRE OR FLAME AND THAT THERE WA NO INJURY AS A RESULT OF THE ISSUE. THIS TYPE OF FAILURE IS TYPICALLY ASSOCIATED WITH DROPPING THE UNIT ONTO A HARD SURFACE OR OTHER TYPE OF SEVERE IMPACT. THE ORIGINAL DESIGN TESTING INVOLVED DROPPING THE UNIT FROM A 1.0 M HEIGHT PER UL260-1 2ND EDITION. PRODUCTION SAMPLES WERE DROPPED THREE TO FIVE TIMES FROM A HEIGHT OF 1.0 M AND THE HOUSING SHOWED DAMAGE AND CRACKING (ONLY AFTER AT LEAST THREE DROPS). THIS PRODUCT WAS RELEASED AS A RESULT OF CAPA (B)(4), WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES FOR WHICH A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING INVESTIGATED IN ORDER TO DETERMINE ROOT CAUSE AND WILL CONTINUE TO BE MONITORED. EVALUATION SUMMARY: A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS CRACKED OPEN, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS PLUGGED IN AND THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED AT 13.87 VDC, WHICH IS NORMAL AND INDICATES THAT THE POWER SUPPLY IS PRODUCING THE CORRECT VOLTAGE. RESISTANCE ACROSS THE DC PLUG WAS MEASURED OPEN, WHICH INDICATES THAT THERE IS NOT A SHOT IN THE DC CORD. SMOKE, FIRE AND SPARKING WERE NOT OBSERVED WHILE THE POWER SUPPLY WAS PLUGGED IN. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 55.7 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW. SEE SCANNED PAGE.
CUSTOMER REPORTED TO CUSTOMER SERVICE THAT THERE WAS A SUCTION ISSUE WITH PUMP ((B)(4)). WHEN RETURNS DEPT RECEIVED THE PUMP, THEY NOTED THAT THE TRANSFORMER HOUSING WAS BROKEN OPEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE | HGX | MEDELA, INC. | 57063/9207010 | REV L-1811 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |