FDA Adverse Event Malfunction Summary report: N

FLEXCATH STEERABLE SHEATH

MDR report key: 2802926 · Received October 24, 2012

Report

Report Number
3002648230-2012-00094
Event Type
Malfunction
Date Received
October 24, 2012
Date of Event
September 27, 2012
Report Date
September 27, 2012
Manufacturer
MEDTRONIC CRYOCATH LP
Product Code
DRA
PMA / PMN Number
K081049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED DEVICE WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. VISUAL INSPECTION SHOWED THAT THE SHAFT WAS INTACT WITH NO APPARENT ISSUES. THE REPORTED ISSUE HAS BEEN CONFIRMED THROUGH TESTING; AIR ASPIRATION WAS REPRODUCED WHEN A CATHETER WAS INTRODUCED THROUGH THE SHEATH. DISSECTION SHOWED THE HEMOSTATIC VALVE WAS LEAKING; VALVE WAS TORN. A FIELD ACTION WAS INITIATED IN JULY 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.

Description of Event or Problem · 1

THE USER REPORTED AIR COMING FROM THE VALVE OF THE SHEATH. NO ADVERSE EVENT OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXCATH STEERABLE SHEATH CATHETER, STEERABLE DRA MEDTRONIC CRYOCATH LP 3FC12

Patients

Seq Age Sex Outcome Treatment
1