FDA Adverse Event
Malfunction
Summary report: N
FLEXCATH STEERABLE SHEATH
MDR report key: 2802926
·
Received October 24, 2012
Report
- Report Number
- 3002648230-2012-00094
- Event Type
- Malfunction
- Date Received
- October 24, 2012
- Date of Event
- September 27, 2012
- Report Date
- September 27, 2012
- Manufacturer
- MEDTRONIC CRYOCATH LP
- Product Code
- DRA
- PMA / PMN Number
- K081049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE RETURNED DEVICE WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. VISUAL INSPECTION SHOWED THAT THE SHAFT WAS INTACT WITH NO APPARENT ISSUES. THE REPORTED ISSUE HAS BEEN CONFIRMED THROUGH TESTING; AIR ASPIRATION WAS REPRODUCED WHEN A CATHETER WAS INTRODUCED THROUGH THE SHEATH. DISSECTION SHOWED THE HEMOSTATIC VALVE WAS LEAKING; VALVE WAS TORN. A FIELD ACTION WAS INITIATED IN JULY 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.
Description of Event or Problem · 1
THE USER REPORTED AIR COMING FROM THE VALVE OF THE SHEATH. NO ADVERSE EVENT OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXCATH STEERABLE SHEATH | CATHETER, STEERABLE | DRA | MEDTRONIC CRYOCATH LP | 3FC12 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |