AFFINITY OXY NT 511
Report
- Report Number
- 2184009-2012-00071
- Event Type
- Malfunction
- Date Received
- October 24, 2012
- Date of Event
- September 5, 2012
- Report Date
- October 1, 2012
- Manufacturer
- PERFUSION SYSTEMS
- Product Code
- DTZ
- PMA / PMN Number
- K932252
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ANALYSIS: UPON RECEIPT AT MEDTRONIC'S QUALITY LABORATORY, VISUAL INSPECTION SHOWED NO OUTWARD SIGNS OF CRACKS OR DAMAGE THROUGHOUT DEVICE OR PORTS. PERFORMANCE ANALYSIS: BLOOD SIDE PRESSURE INTEGRITY TESTING WAS PERFORMED AT THREE LITERS PER MINUTE WITH TWENTY-THREE PSI OF BACK PRESSURE FOR TEN MINUTES. DURING THE TEST A LEAK FROM THE HEX SIDE SEAM WAS OBSERVED. THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS PRODUCT MET ALL MANUFACTURING SPECIFICATION FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT. CONCLUSION: BASED ON ANALYSIS AND INVESTIGATION, THE LEAK WAS CONFIRMED TO ORIGINATE FROM THE HEX SIDE SEAM. A PARTIAL VOID MAY HAVE FORMED DURING ADHESIVE DISPENSING INTO THE ADHESIVE GROOVE OF THE HEAT EXCHANGER ALLOWING FOR ACCEPTABLE PRESSURE TEST RESULTS PRIOR TO DISTRIBUTION. IT IS POSSIBLE A PHYSICAL SHOCK ENCOUNTERED DURING SHIPPING OR HANDLING OF THE PRODUCT MAY HAVE COMPROMISED THAT BOND. HISTORICALLY, IT HAS BEEN OBSERVED THAT OVERLY AGGRESSIVE SHIPPING AND HANDLING CAN CAUSE THE PARTIAL BOND TO BECOME COMPROMISED RESULTING IN LEAK AND CONTRIBUTED TO THE EVENT. MEDTRONIC IS MONITORING FOR SIMILAR COMPLAINTS. (B)(6). (B)(4).
MEDTRONIC RECEIVED INFORMATION THAT DURING THE PRIMING PHASE OF THIS OXYGENATOR, A LEAK WAS OBSERVED TO BE COMING FROM BETWEEN THE HEAT EXCHANGER AND FIBER BUNDLE. THE DEVICE WAS CHANGED OUT WITH NO PATIENT EFFECT REPORTED AND RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY OXY NT 511 | OXYGENATOR, CARDIOPULMONARY BYPASS | DTZ | PERFUSION SYSTEMS | 95212 | 11656186 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |