FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC CLIP APPLIER

MDR report key: 2802799 · Received October 24, 2012

Report

Report Number
3005075853-2012-04844
Event Type
Malfunction
Date Received
October 24, 2012
Date of Event
October 23, 2012
Report Date
October 23, 2012
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION REQUESTED: WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? UNK. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? UNK. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? UNK. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? UNK. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? UNK. WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? UNK. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? UNK. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? UNK.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED AND FED AND FORMED TWO SCISSORED CLIPS, AND THEN THE REMAINING CLIPS WERE FORMED AS INTENDED. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED. ALTHOUGH IT IS NOT POSSIBLE TO CONCLUDE HOW THE CIRCUMSTANCES OCCURRED, IT IS KNOWN FROM THE HISTORY OF THE INSTRUMENT THAT AN INCORRECT/EXCESSIVE APPLICATION OF TORQUE TO THE JAWS DURING INSTRUMENT USE CREATES A MISALIGNMENT OF THE TIPS. IN ADDITION, AS PER THE INSTRUCTIONS FOR USE "DO NOT EXCESSIVELY TWIST OR TORQUE THE INSTRUMENT JAWS WHEN POSITIONING THE INSTRUMENT ON A VESSEL AND FIRING. EXCESSIVE TWISTING OR TORQUING MAY RESULT IN CLIP MALFORMATION". THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIPS WOULD NOT CLOSE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT REPORTED. NO OTHER DETAILS OF THE EVENT WERE PROVIDED.