LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2012-04844
- Event Type
- Malfunction
- Date Received
- October 24, 2012
- Date of Event
- October 23, 2012
- Report Date
- October 23, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION REQUESTED: WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? UNK. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? UNK. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? UNK. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? UNK. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? UNK. WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? UNK. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? UNK. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? UNK.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED AND FED AND FORMED TWO SCISSORED CLIPS, AND THEN THE REMAINING CLIPS WERE FORMED AS INTENDED. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED. ALTHOUGH IT IS NOT POSSIBLE TO CONCLUDE HOW THE CIRCUMSTANCES OCCURRED, IT IS KNOWN FROM THE HISTORY OF THE INSTRUMENT THAT AN INCORRECT/EXCESSIVE APPLICATION OF TORQUE TO THE JAWS DURING INSTRUMENT USE CREATES A MISALIGNMENT OF THE TIPS. IN ADDITION, AS PER THE INSTRUCTIONS FOR USE "DO NOT EXCESSIVELY TWIST OR TORQUE THE INSTRUMENT JAWS WHEN POSITIONING THE INSTRUMENT ON A VESSEL AND FIRING. EXCESSIVE TWISTING OR TORQUING MAY RESULT IN CLIP MALFORMATION". THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIPS WOULD NOT CLOSE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT REPORTED. NO OTHER DETAILS OF THE EVENT WERE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | J4C46Y |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |