DIMENSION(R) CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2012-00289
- Event Type
- Malfunction
- Date Received
- October 24, 2012
- Date of Event
- September 23, 2012
- Report Date
- September 24, 2012
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- LCP
- PMA / PMN Number
- K102510
- Removal / Correction Number
- 2517506-08-29-2012-013-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE (B)(6) RESULTS IS USER ERROR. THE ACCOUNT FAILED TO INPUT THE CORRECT LOT SPECIFIC SCALERS FOR THE LOT OF (B)(6) REAGENT. THE FALSELY ELEVATED RESULTS WERE CAUSED BY USING INCORRECT SCALERS AND THE ISSUE WAS RESOLVED BY ENTERING CORRECT SCALERS AND VERIFYING WITH QC. THE (B)(6) METHOD USES AN ADDITIONAL PARAMETER CALLED SCALER VALUES. THESE VALUES ARE POLYNOMIAL EQUATION FACTORS THAT HAVE BEEN DETERMINED FOR (B)(6) IN ORDER TO PROVIDE THE BEST CORRELATION TO THE (B)(6) REFERENCE METHODOLOGY. THIS FEATURE WAS COMMUNICATED TO CUSTOMERS WITH THE LAUNCH OF (B)(6) ((B)(4)) IN THE (B)(6) KIT SUPPLEMENT. SIEMENS HEALTHCARE DIAGNOSTICS INC. ISSUED AN URGENT MEDICAL DEVICE CORRECTION IN AUGUST 2012 TO REINFORCE INSTRUCTIONS TO CUSTOMERS TO ENTER AND VERIFY SCALERS WITH EVERY CALIBRATION OF NEW (B)(6) FLEX(R) REAGENT CARTRIDGE LOTS AND WITH EACH RECALIBRATION OF THE SAME FLEX(R) LOT. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A GROUP OF (B)(6) RESULTS WAS OBTAINED ON QC AND PATIENT SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE RESULTS WERE EVENTUALLY QUESTIONED AND IT WAS DETERMINED THAT INCORRECT METHOD SCALERS HAD BEEN IN USE FOR A PERIOD IN WHICH AT LEAST 6 PATIENT RESULTS HAD BEEN REPORTED. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE (B)(6) RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION(R) CLINICAL CHEMISTRY SYSTEM | HEMOGLOBIN A1C KIT | LCP | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | GA3134 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |