FDA Adverse Event Injury Summary report: N

PINNACLE MTL INS NEUT36IDX58OD

MDR report key: 2802758 · Received October 24, 2012

Report

Report Number
1818910-2012-26626
Event Type
Injury
Date Received
October 24, 2012
Date of Event
October 16, 2012
Report Date
March 8, 2013
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
KWA
PMA / PMN Number
K083642
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS SINCE THEIR RELEASE FOR DISTRIBUTION. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED.DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

**UPDATE** 3/8/2013- LITIGATION PAPERS RECEIVED. IT IS ALLEGED THAT THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, GRINDING SENSATION, AUDIBLE SQUEAKING, AND TOXIC AMOUNTS OF COBALT CHROMIUM METAL DEBRIS TO BE RELEASED INTO THE PATIENTS TISSUE. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE INVESTIGATION. THE INVESTIGATION IS ONGOING. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE FOUND NO ADDITIONAL RELATED REPORTS FOR THE REPORTED PART AND LOT CODE COMBINATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

THE PATIENT WAS REVISED BECAUSE OF PAIN. (LEFT HIP).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PINNACLE MTL INS NEUT36IDX58OD HIP LINER KWA DEPUY ORTHOPAEDICS INC US 3111793

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention