FDA Adverse Event
Summary report: N
840 VENTILATOR
MDR report key: 2802626
·
Received October 19, 2012
Report
- Report Number
- 8020893-2012-00980
- Date Received
- October 19, 2012
- Date of Event
- September 26, 2012
- Report Date
- October 2, 2012
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION IN THE MEMORY BUT THE CSE WAS NOT ABLE TO DUPLICATE THE PROBLEM. THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) CPU PCB AS PRECAUTIONARY ACTION. THE UNIT PASSED EXTENDED SELF TESTING.