FDA Adverse Event Summary report: N

840 VENTILATOR

MDR report key: 2802626 · Received October 19, 2012

Report

Report Number
8020893-2012-00980
Date Received
October 19, 2012
Date of Event
September 26, 2012
Report Date
October 2, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION IN THE MEMORY BUT THE CSE WAS NOT ABLE TO DUPLICATE THE PROBLEM. THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) CPU PCB AS PRECAUTIONARY ACTION. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1