FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2802538 · Received October 19, 2012

Report

Report Number
1720753-2012-08447
Event Type
Malfunction
Date Received
October 19, 2012
Date of Event
October 8, 2012
Report Date
October 19, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INTERCONNECT CABLE AND THE POWER SUPPLY WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY LOCKED UP. THERE IS NO REPORT OF PT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1