FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE

MDR report key: 2802497 · Received October 18, 2012

Report

Report Number
1419937-2012-00644
Event Type
Malfunction
Date Received
October 18, 2012
Date of Event
September 21, 2012
Report Date
September 21, 2012
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. IN FOLLOW UP WITH THE CUSTOMER, HE INDICATED THAT THE POWER SUPPLY CASING FELL APART WHEN UNPLUGGING IT FROM THE WALL. INDICATED THAT THE POWER SUPPLY IS NOT ALWAYS UNPLUGGED AFTER USE, AND THAT HE STRAPPED IT BACK TOGETHER TO KEEP USING. CUSTOMER STATED THERE WAS NO SPARK, FIRE, FLAME, OR INJURY RECEIVED. A PRODUCT EVALUATION REVEALED THAT THE TRANSFORMER HOUSING WAS BROKEN OPEN, WHICH IS A SAFETY RISK. THIS TYPE OF FAILURE IS TYPICALLY ASSOCIATED WITH DROPPING THE UNIT ONTO A HARD SURFACE OR OTHER TYPE OF SEVERE IMPACT. THE ORIGINAL DESIGN TESTING INVOLVED DROPPING THE UNIT FROM A 1.0 M HEIGHT PER UL2601-1 2ND EDITION. PRODUCTION SAMPLES WERE DROPPED THREE TO FIVE TIMES FROM A HEIGHT OF 1.0 M AND THE HOUSINGS SHOWED DAMAGE AND CRACKING (ONLY AFTER AT LEAST THREE DROPS). THIS PRODUCT WAS RELEASED AS A RESULT OF CAPA (B)(4) WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES FOR WHICH A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING INVESTIGATED IN ORDER TO DETERMINE ROOT CAUSE AND WILL CONTINUE TO BE MONITORED. EVALUATION SUMMARY: A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS CRACKED OPEN. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS PLUGGED IN AND THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED AT 13.9 VDC, WHICH IS NORMAL AND INDICATES THAT THE POWER SUPPLY IS PRODUCING THE CORRECT VOLTAGE. RESISTANCE ACROSS THE DC PLUG WAS MEASURED OPEN, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. SMOKE, FIRE AND SPARKING WERE NOT OBSERVED WHILE THE POWER SUPPLY WAS PLUGGED IN. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 55.8 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT THE POWER SUPPLY FOR THEIR BREAST PUMP FELL APART WHEN THEY DISCONNECTED IT FROM THE OUTLET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE HGX MEDELA, INC. 57063/9207010 REV L 2011

Patients

Seq Age Sex Outcome Treatment
1