COMFORTDRIVE 200XDR
Report
- Report Number
- 3003637274-2012-00051
- Event Type
- Malfunction
- Date Received
- October 18, 2012
- Date of Event
- September 17, 2012
- Report Date
- September 24, 2012
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EIA
- PMA / PMN Number
- K080677
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- DENTIST
Narratives
THE ANALYSIS OF THE HANDPIECE DID SHOW THAT THE HEAD HAD A DENT. THIS CAUSED THE BACK CAP BUTTON TO STICK IN, CAUSING UNUSUAL INNER FRICTION, CAUSING FINALLY THE HEAD TO HEAT UP. ROOT CAUSE IS OUT OF EXPERIENCE A STRONG HIT, E.G. A DROP DURING THE REPROCESSING. ALSO THE BEARINGS HAVE BEEN GRITTY. THIS MIGHT BE THE RESULT OF A STRONGER WEAR PROCESS DUE TO THE HIGH TEMPERATURES. THE USER INSTRUCTION REQUIRES A VISUAL AND FUNCTIONAL TEST PRIOR TO EACH TREATMENT TO ENSURE PRODUCT IS WITHIN SPECIFICATION, WHICH WOULD HENCE BE MANDATORY. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.
DURING A STANDARD DENTAL TREATMENT, THE HANDPIECE GOT HOT AND BURNED THE PT ON CHEEK APPROXIMATELY TO THE SIZE OF THE HANDPIECE HEAD. PT WAS GIVEN SOME SAMPLES OF ORABASE OINTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COMFORTDRIVE 200XDR | DENTAL HANDPIECE | EIA | KAVO DENTAL GMBH | 200XDR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |