FDA Adverse Event Malfunction Summary report: N

COMFORTDRIVE 200XDR

MDR report key: 2802473 · Received October 18, 2012

Report

Report Number
3003637274-2012-00051
Event Type
Malfunction
Date Received
October 18, 2012
Date of Event
September 17, 2012
Report Date
September 24, 2012
Manufacturer
KAVO DENTAL GMBH
Product Code
EIA
PMA / PMN Number
K080677
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS OF THE HANDPIECE DID SHOW THAT THE HEAD HAD A DENT. THIS CAUSED THE BACK CAP BUTTON TO STICK IN, CAUSING UNUSUAL INNER FRICTION, CAUSING FINALLY THE HEAD TO HEAT UP. ROOT CAUSE IS OUT OF EXPERIENCE A STRONG HIT, E.G. A DROP DURING THE REPROCESSING. ALSO THE BEARINGS HAVE BEEN GRITTY. THIS MIGHT BE THE RESULT OF A STRONGER WEAR PROCESS DUE TO THE HIGH TEMPERATURES. THE USER INSTRUCTION REQUIRES A VISUAL AND FUNCTIONAL TEST PRIOR TO EACH TREATMENT TO ENSURE PRODUCT IS WITHIN SPECIFICATION, WHICH WOULD HENCE BE MANDATORY. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.

Description of Event or Problem · 1

DURING A STANDARD DENTAL TREATMENT, THE HANDPIECE GOT HOT AND BURNED THE PT ON CHEEK APPROXIMATELY TO THE SIZE OF THE HANDPIECE HEAD. PT WAS GIVEN SOME SAMPLES OF ORABASE OINTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMFORTDRIVE 200XDR DENTAL HANDPIECE EIA KAVO DENTAL GMBH 200XDR

Patients

Seq Age Sex Outcome Treatment
1 56 YR