COULTER® ACT DIFF 2¿ ANALYZER
Report
- Report Number
- 1061932-2012-02610
- Event Type
- Malfunction
- Date Received
- October 23, 2012
- Date of Event
- September 18, 2012
- Report Date
- September 28, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K990352
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
THE FIELD SERVICE ENGINEER (FSE) REPLACED THE ASPIRATION PUMP AND RESOLVED THE ISSUE. THE FSE MANUALLY ADJUSTED THE RED BLOOD CELL (RBC), WHITE BLOOD CELL (WBC), HEMOGLOBIN (HGB) AND PLATELET (PLT) CALIBRATION FACTORS TO CONTROLS. THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. ALL RELATED MEDWATCH REPORTS ASSOCIATED WITH THIS EVENT: 1061932-2012-02608, 1061932-2012-02609, 1061932-2012-02610, 1061932-2012-02611, 1061932-2012-02612.
THE CUSTOMER REPORTED LOW WHITE BLOOD CELL (WBC), RED BLOOD CELL (RBC), HEMOGLOBIN (HGB), HEMATOCRIT (HCT), AND PLATELET (PLT) RESULTS, FOR FIVE PATIENTS, ON SEPARATE DAYS INVOLVING THE COULTER ACT DIFF 2 ANALYZER. THIS IS REPORT THREE OF FIVE. THE REANALYZED RESULTS CORRELATED TO THE CUSTOMER'S EXPECTATION AND WERE CONSIDERED CORRECT. THE ERRONEOUS RESULTS WERE NOT RELEASED OUT OF THE LABORATORY. THERE WAS NO PATIENT CONSEQUENCE OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® ACT DIFF 2¿ ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |