FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2802159 · Received October 23, 2012

Report

Report Number
3004209178-2012-90317
Event Type
Malfunction
Date Received
October 23, 2012
Date of Event
September 15, 2012
Report Date
October 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED AN ERROR DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN ALARMED AN ERROR AND INSULIN WAS SQUIRTING OUT DURING THE MANUAL PRIME PROCESS. THE BLOOD GLUCOSE READING WAS 83MG/DL. ADVISED THE CUSTOMER TO DISCONNECT FROM THE INSULIN PUMP. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 23 YR