FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2802157
·
Received October 23, 2012
Report
- Report Number
- 3004209178-2012-90314
- Event Type
- Malfunction
- Date Received
- October 23, 2012
- Date of Event
- September 14, 2012
- Report Date
- October 17, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD CRACKED CASE ON THE DISPLAY WINDOW.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT INSULIN SQUIRTED OUT AND THE NUMBERS DID NOT APPEAR ON THE SCREEN WHILE PRIMING THE DEVICE. THE CUSTOMER CHANGED THE INFUSION SET TWICE WITHOUT RESULTS. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE INSULIN PUMP. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |