FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2802157 · Received October 23, 2012

Report

Report Number
3004209178-2012-90314
Event Type
Malfunction
Date Received
October 23, 2012
Date of Event
September 14, 2012
Report Date
October 17, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD CRACKED CASE ON THE DISPLAY WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN SQUIRTED OUT AND THE NUMBERS DID NOT APPEAR ON THE SCREEN WHILE PRIMING THE DEVICE. THE CUSTOMER CHANGED THE INFUSION SET TWICE WITHOUT RESULTS. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE INSULIN PUMP. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 62 YR