FDA Adverse Event Malfunction Summary report: N

OT VERIO PRO METER

MDR report key: 2802137 · Received October 23, 2012

Report

Report Number
3008382007-2012-05469
Event Type
Malfunction
Date Received
October 23, 2012
Report Date
September 28, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2012, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) IN (B)(4) ALLEGING A ONETOUCH VERIO PRO METER WAS DISPLAYING AN ERROR 4 MESSAGE. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY DUE TO THE ALLEGED ISSUE. THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION(S): THERE IS NO INDICATION THAT SUBJECT METER CAUSE OR CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT/REPORTER DID NOT REPORT ANY BLOOD GLUCOSE READINGS, SYMPTOMS OR TREATMENT SUGGESTIVE THAT AN ACUTE COMPLICATION OF DIABETES OCCURRED. HOWEVER THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT VERIO PRO METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3237274

Patients

Seq Age Sex Outcome Treatment
1