FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2802136
·
Received October 23, 2012
Report
- Report Number
- 3004209178-2012-90309
- Event Type
- Malfunction
- Date Received
- October 23, 2012
- Date of Event
- September 12, 2012
- Report Date
- October 15, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD MISSING END CAP STICKER.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT HER INSULIN PUMP WAS STUCK IN THE PRIME/REWIND LOOP, AND INSULIN WAS SQUIRTING OUT. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT WHEN SHE HOLDS DOWN THE ACTIVATE BUTTON SHE DOES NOT RECEIVE BEEPS OR NUMBERS. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |