FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2802114
·
Received October 23, 2012
Report
- Report Number
- 3004209178-2012-90305
- Event Type
- Malfunction
- Date Received
- October 23, 2012
- Date of Event
- September 3, 2012
- Report Date
- October 1, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED AN ERROR DURING THE BASIC OCCLUSION TEST, AND THE DEVICE WAS UNABLE TO PRIME DURING REWIND AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD AN ERROR ALARM DURING THE MANUAL PRIME PROCESS. ADVISED THE CUSTOMER THAT THE DEVICE WOULD BE REPLACED. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 290MG/DL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |