FDA Adverse Event Death Summary report: N

ATTAIN STARFIX

MDR report key: 2802068 · Received October 23, 2012

Report

Report Number
2649622-2012-15612
Event Type
Death
Date Received
October 23, 2012
Date of Event
January 8, 2009
Report Date
December 3, 2012
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P060039
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED, AND ANALYSIS RESULTS REVEALED NO ANOMALIES FOUND. IT WAS FURTHER NOTED THAT THERE WAS BLOOD ON THE CONDUCTOR (NOT OBSTRUCTED), THE CONDUCTOR WAS DISTORTED, AND THERE WAS APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND ANALYSIS RESULTS REVEALED NO ANOMALIES FOUND.

Description of Event or Problem · 1

THE SYSTEM WAS RETURNED WITH NO INFORMATION. A DATABASE SEARCH BY SERIAL NUMBER SHOWED THE PATIENT EXPIRED LESS THAN 1 YEAR AFTER IMPLANT. THE DEATH OCCURRED GREAT THAN 2 YEARS AGO. NO COMPLAINTS, ALLEGATIONS, OR, PREVIOUS CONTACTS, REGARDING THE SYSTEM OR ANY INDIVIDUAL COMPONENTS HAS BEEN RECEIVED.

Description of Event or Problem · 1

THE SYSTEM WAS RETURNED WITH NO INFORMATION. A DATABASE SEARCH BY SERIAL NUMBER SHOWED THE PATIENT EXPIRED LESS THAN 1 YEAR AFTER IMPLANT. THE DEATH OCCURRED GREAT THAN 2 YEARS AGO. NO COMPLAINTS, ALLEGATIONS, OR, PREVIOUS CONTACTS, REGARDING THE SYSTEM OR ANY INDIVIDUAL COMPONENTS HAS BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN STARFIX DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 4195

Patients

Seq Age Sex Outcome Treatment
1 77 YR Death LEAD/STJ IMPLANTABLE PACING LEAD| LEAD/STJ IMPLANTABLE PACING LEAD