FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2802002 · Received October 23, 2012

Report

Report Number
2032227-2012-07522
Event Type
Malfunction
Date Received
October 23, 2012
Date of Event
October 12, 2012
Report Date
October 12, 2012
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A LOOSE DRIVE SUPPORT DISK. NO ALARMS WERE NOTED. THE INSULIN PUMP HAD A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, A CRACKED DISPLAY WINDOW, A SCRATCHED DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS AND A BROKEN RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN ALARM DURING THE MANUAL PRIME. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC MINIMED MMT-722RNAP

Patients

Seq Age Sex Outcome Treatment
1 29 YR