FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2802002
·
Received October 23, 2012
Report
- Report Number
- 2032227-2012-07522
- Event Type
- Malfunction
- Date Received
- October 23, 2012
- Date of Event
- October 12, 2012
- Report Date
- October 12, 2012
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A LOOSE DRIVE SUPPORT DISK. NO ALARMS WERE NOTED. THE INSULIN PUMP HAD A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, A CRACKED DISPLAY WINDOW, A SCRATCHED DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS AND A BROKEN RESERVOIR TUBE LIP.
Description of Event or Problem · 1
THE CUSTOMER REPORTED AN ALARM DURING THE MANUAL PRIME. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC MINIMED | MMT-722RNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |