FDA Adverse Event Malfunction Summary report: N

ADVANTAGE BALLOON CATHETER INFLATION DEVICE

MDR report key: 2801926 · Received October 23, 2012

Report

Report Number
2134265-2012-06511
Event Type
Malfunction
Date Received
October 23, 2012
Date of Event
September 24, 2012
Report Date
September 24, 2012
Manufacturer
BOSTON SCIENTIFIC - CORK
Product Code
MAV
PMA / PMN Number
K955869
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, AN ISSUE OCCURRED WITH THE PRESSURE GAUGE. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE NON-CALCIFIED AND MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING ARTERY. THE PRESSURE GAUGE DID NOT GO UP WHILE INFLATING AN UNSPECIFIED DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.