FDA Adverse Event Malfunction Summary report: N

RX ACCULINK CAROTID STENT SYSTEM

MDR report key: 2801893 · Received October 23, 2012

Report

Report Number
2024168-2012-06690
Event Type
Malfunction
Date Received
October 23, 2012
Date of Event
October 2, 2012
Report Date
October 3, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
NIM
PMA / PMN Number
P040012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE PREMATURE DEPLOYMENT WAS ABLE TO BE CONFIRMED. THE RESISTANCE WITH THE SHEATH COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON A VISUAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING INSERTION OF THE ACCULINK DEVICE INTO A 7 FRENCH NON-ABBOTT SHEATH, SLIGHT RESISTANCE WAS ENCOUNTERED AND 3 MM OF STENT STRUTS BECAME EXPOSED DESPITE THE HANDLE BEING LOCKED. THE ACCULINK WAS EXCHANGED FOR AN XACT 7-9 X 40 MM STENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THERE WAS NO CLINICALLY SIGNIFICANT DELAY OF PROCEDURE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RX ACCULINK CAROTID STENT SYSTEM CAROTID STENT SYSTEM NIM AV-TEMECULA-CT 1051161

Patients

Seq Age Sex Outcome Treatment
1 SHEATH: COOK 7F SHUTTLE SHEATH