FDA Adverse Event
Malfunction
Summary report: N
MILLENNIUM INFANT VENTILATOR
MDR report key: 2801578
·
Received October 18, 2012
Report
- Report Number
- 2020676-2012-00043
- Event Type
- Malfunction
- Date Received
- October 18, 2012
- Date of Event
- September 20, 2012
- Report Date
- October 18, 2012
- Manufacturer
- SECHRIST INDUSTRIES, INC.
- Product Code
- CBK
- PMA / PMN Number
- K993167
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DISTRIBUTOR AND NOTIFIER ADDRESS: (B)(6).
Description of Event or Problem · 1
DURING A PRE-CLINICAL CHECK, AFTER THE VENTILATOR WAS TURNED ON, THE SCREEN WOULD NOT DISPLAY ANY OF THE PARAMETERS. THERE WERE NO VISUAL OR AUDIBLE ALARMS. THE VENTILATOR WAS TAKEN OUT OF SERVICE. THE DISTRIBUTOR REPLACED THE INTERFACE BOARD AND THE ISSUE WAS RESOLVED. THE INTERFACE BOARD IS TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. A FOLLOW-UP SUBMISSION WILL BE SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MILLENNIUM INFANT VENTILATOR | MILLENNIUM VENTILATOR | CBK | SECHRIST INDUSTRIES, INC. | 23065-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |