FDA Adverse Event Malfunction Summary report: N

MILLENNIUM INFANT VENTILATOR

MDR report key: 2801578 · Received October 18, 2012

Report

Report Number
2020676-2012-00043
Event Type
Malfunction
Date Received
October 18, 2012
Date of Event
September 20, 2012
Report Date
October 18, 2012
Manufacturer
SECHRIST INDUSTRIES, INC.
Product Code
CBK
PMA / PMN Number
K993167
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DISTRIBUTOR AND NOTIFIER ADDRESS: (B)(6).

Description of Event or Problem · 1

DURING A PRE-CLINICAL CHECK, AFTER THE VENTILATOR WAS TURNED ON, THE SCREEN WOULD NOT DISPLAY ANY OF THE PARAMETERS. THERE WERE NO VISUAL OR AUDIBLE ALARMS. THE VENTILATOR WAS TAKEN OUT OF SERVICE. THE DISTRIBUTOR REPLACED THE INTERFACE BOARD AND THE ISSUE WAS RESOLVED. THE INTERFACE BOARD IS TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. A FOLLOW-UP SUBMISSION WILL BE SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MILLENNIUM INFANT VENTILATOR MILLENNIUM VENTILATOR CBK SECHRIST INDUSTRIES, INC. 23065-1

Patients

Seq Age Sex Outcome Treatment
1