FDA Adverse Event Malfunction Summary report: N

VIPER2 X-TAB POLYAXIAL DRIVER

MDR report key: 2801538 · Received October 23, 2012

Report

Report Number
1526439-2012-00223
Event Type
Malfunction
Date Received
October 23, 2012
Report Date
April 22, 2013
Manufacturer
DEPUY RAYNHAM
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEPUY SPINE HAS REQUESTED RETURN OF THE DRIVER FOR EVALUATION. A FOLLOW UP REPORT WILL BE FILED UPON RECEIPT OF THE DEVICE AND COMPLETION OF THE EVALUATION.

Additional Manufacturer Narrative · 1

THE VIPER2 X-TAB POLYAXIAL DRIVER IS NOT AVAILABLE FOR EVALUATION. REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DISCREPANCIES AND CONFIRMED THAT THE PRODUCT WAS RELEASED ACCOMPLISHING ALL QUALITY REQUIREMENTS. WITHOUT A PRODUCT SAMPLE, WE ARE UNABLE TO CONFIRM THE REPORTED ISSUE OR IDENTIFY THE ROOT CAUSE. HOWEVER, IT SHOULD BE NOTED THAT A CAPA WAS OPENED TO ADDRESS PREVIOUS PRODUCT COMPLAINTS INVOLVING DRIVER TIP BREAKAGE. THIS PRODUCTION LOT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THE CAPA. IN THE ABSENCE OF A PRODUCT SAMPLE, THIS COMPLAINT WILL BE CLOSED WITH NO FURTHER ACTION REQUIRED. IF THE COMPLAINT PRODUCT SAMPLE BECOMES AVAILABLE, THE COMPLAINT FILE WILL BE RE-OPENED AND THE PRODUCT EVALUATED.

Additional Manufacturer Narrative · 1

VISUAL EXAMINATION OF THE RETURNED DRIVER FOUND THAT THE MOST DISTAL PORTION OF THE TIP HAD BROKEN OFF FROM THE INSTRUMENT. REVIEW OF DEVICE HISTORY RECORDS CONFIRMED THE INSTRUMENT WAS MANUFACTURED TO SPECIFICATION REQUIREMENTS. A CAPA THAT ADDRESSED TIP BREAKAGE THROUGH DESIGN MODIFICATION AND MATERIAL CHANGE HAS BEEN IMPLEMENTED. THIS PRODUCT LOT WAS MANUFACTURED PRIOR TO THOSE CHANGES. AT THIS TIME, THE COMPLAINT IS CONSIDERED TO BE CLOSED.

Description of Event or Problem · 1

INTERNATIONAL AFFILIATE REPORTS THE INSPECTION OF A RETURNED LOANER KIT FOUND THE TIP OF THE DRIVER HAD BROKEN OFF FROM THE INSTRUMENT. THE BROKEN TIP WAS NOT PRESENT WITHIN THE KIT. IT IS NOT KNOWN WHEN OR WHERE TIP BREAKAGE OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIPER2 X-TAB POLYAXIAL DRIVER INSTRUMENT, ORTHOAPEDIC, MANUAL, SURGICAL LXH DEPUY RAYNHAM MI19233

Patients

Seq Age Sex Outcome Treatment
1