ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM

MDR report key: 2801410 · Received October 18, 2012

Report

Report Number
2028159-2012-01621
Event Type
Malfunction
Date Received
October 18, 2012
Date of Event
September 18, 2012
Report Date
September 18, 2012
Manufacturer
ALCON - IRVINE TECHNOLOGY CTR
Product Code
MPA
PMA / PMN Number
K951627
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP EXAMINED THE SYS AND REPLACED THE XENON LAMP. PREVENTIVE MAINTENANCE WAS PERFORMED. THE SYS WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVAL AND NO ADDITIONAL INFO PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID INDICATE ONE SIMILAR REPORT FOR THIS SYSTEM. SINCE NO SAMPLE HAS BEEN RETURNED FOR EVAL THE ROOT CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED CONCLUSIVELY. HOWEVER, THE MOST LIKELY CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO A NONCONFORMING XENON LAMP. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT DURING SURGERY, A SYSTEM MESSAGE DISPLAYED AND THE LAMP WAS OVERLY WARM. THE SYSTEM WAS SWITCHED OUT TO COMPLETE THE CASE. THERE WAS NO HARM TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM ENDOILLUMINATOR MPA ALCON - IRVINE TECHNOLOGY CTR NA NA