PWRD ECH FLEX 60MM
Report
- Report Number
- 3005075853-2012-04821
- Event Type
- Malfunction
- Date Received
- October 23, 2012
- Date of Event
- August 6, 2012
- Report Date
- August 6, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DAMAGED RELEASE BUTTON. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING WAS OBTAINED: ON WHAT TISSUE TYPE WAS THE DEVICE USED? STOMACH. AT WHAT LOCATION ON THE TISSUE? ATRIUM TO FUNDUS 1ST AND 2ND FIRINGS. ON WHICH FIRING(S) DID THIS EVENT OCCUR (1ST, 2ND, 12TH, ETC.)? 3RD. WHAT COLOR CARTRIDGE WAS BEING USED? BLUE. WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? UNK. WAS BUTTRESSING MATERIAL UTILIZED? NA. WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP? UNK. WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? WERE ANY OF THE FORCES HIGHER OR LOWER THAN EXPECTED (CLOSING, FIRING, OR OPENING)? NA. AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? NA. WAS THERE ANY DIFFICULTY REMOVING THE DEVICE FROM THE TISSUE? NA. IS THERE ANY OTHER ADDITIONAL INFORMATION YOU WOULD LIKE TO SHARE ABOUT THIS DEVICE OR PROCEDURE? DEVICE WAS BEING LOADED AND WHEN CLOSED THE JAWS WOULD POP OPEN. WHAT WERE THE PATIENT¿S PRE-EXISTING CONDITIONS? OBESITY. DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? IF SO, WHAT? UNK. HOW LONG HAS THE SURGEON BEEN USING THIS DEVICE FOR THIS PROCEDURE (IN YEARS)? UNK. WAS THERE A RECENT CONVERSION TO EES DEVICES IN THIS ACCOUNT OR WITH THIS SURGEON? UNK. THE ANALYSIS FOUND THAT ONE DEVICE WAS RECEIVED FOR ANALYSIS WITHOUT CARTRIDGE RELOAD. FURTHERMORE, THE DEVICE WAS FOUND TO HAVE THE CLAMPING MECHANISM DAMAGED. IN ORDER TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS, THE DEVICE WAS DISASSEMBLED. UPON DISASSEMBLING, THE CLAMP ARM WAS NOTED TO BE WORN OUT, IT APPEARS THAT THE DEVICE WAS FORCED OPEN WITH THE KNIFE WAS NOT ON THE HOME POSITION. THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT THE DEVICE WAS USED DURING A SLEEVE GASTRECTOMY. AFTER SUCCESSFULLY FIRING THE DEVICE TWICE. WHEN THE SCRUB NURSE WAS RELOADING THE DEVICE FOR THE 3RD FIRING, WHEN SHE CLOSED THE DEVICE IT POPPED BACK OPEN AND WOULD NOT CLOSE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PWRD ECH FLEX 60MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | J4AW2X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ECR60B |