FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2801260 · Received October 17, 2012

Report

Report Number
8020893-2012-00961
Event Type
Malfunction
Date Received
October 17, 2012
Date of Event
October 15, 2012
Report Date
October 15, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR STOPPED CYCLING. NO PT INVOLVEMENT. THE CUSTOMER REPORTED TO HAVE INSPECTED THE DEVICE AND COULD NOT DUPLICATE THE REPORTED FAILURE. NO PARTS WERE REPLACED. THE UNIT PASSED EXTENDED SELF TESTING. COVIDIEN WAS NOT AUTHORIZED TO SERVICE THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1