FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE

MDR report key: 2801124 · Received October 17, 2012

Report

Report Number
1419937-2012-00642
Event Type
Malfunction
Date Received
October 17, 2012
Date of Event
September 18, 2012
Report Date
September 18, 2012
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. IN F/U WITH THE CUSTOMER, THE CUSTOMER INDICATED THAT THEIR TRANSFORMER HAD FALLEN APART AND THE UNDERLYING ELECTRONICS WERE EXPOSED. A PRODUCT EVAL REVEALED THAT THE TRANSFORMER HOUSING WAS BROKEN, EXPOSING INNER ELECTRONICS, WHICH IS SAFETY RISK. THIS TYPE OF FAILURE IS TYPICALLY ASSOCIATED WITH DROPPING THE UNIT ONTO A HARD SURFACE OR OTHER TYPE OF SEVERE IMPACT. THE ORIGINAL DESIGN TESTING INVOLVED DROPPING THE UNIT FROM A 1.0 M HEIGHT PER UL2601-1 2ND EDITION. PRODUCTION SAMPLES WERE DROPPED THREE TO FIVE TIMES FROM A HEIGHT OF 1.0 M AND THE HOUSINGS SHOWED DAMAGE AND CRACKING (ONLY AFTER AT LEAST THREE DROPS). THIS PRODUCT WAS RELEASED AS A RESULT OF CAPA (B)(4) WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES FOR WHICH A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING INVESTIGATED IN ORDER TO DETERMINE ROOT CAUSE AND WILL CONTINUE TO BE MONITORED. EVAL SUMMARY: A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS CRACKED OPEN, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS PLUGGED IN AND THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED AT 14 VDC, WHICH IS NORMAL AND INDICATES THAT THE POWER SUPPLY IS PRODUCING THE CORRECT VOLTAGE. RESISTANCE ACROSS THE DC PLUG WAS MEASURED OPEN, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. SMOKE, FIRE AND SPARKING WERE NOT OBSERVED WHILE THE POWER SUPPLY WAS PLUGGED IN. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 53.83 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY SMELLED AN ELECTRICAL ODOR. WHEN THEY WERE CONTACTED FOR F/U THEY INDICATED THAT THE TRANSFORMER HAD FALLEN APART, EXPOSING THE UNDERLYING ELECTRONICS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE HGX MEDELA, INC. 57063 / 9207010 REV L - 1511