ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-12620
- Event Type
- Malfunction
- Date Received
- October 23, 2012
- Report Date
- September 26, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED AND WAS EVALUATED ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: REVIEW OF THE BLACK BOX AND DOWNLOAD HISTORY REVEALED SEVERAL LOSS OF PRIME WARNINGS RECORDED IN THE BLACK BOX AS WELL AS SEVERAL TIME AND DATE RESETS. THE BATTERY CAP THAT WAS RETURNED WITH THE PUMP FOR INVESTIGATION HAD CORROSION ON THE SPRING. ON EXAMINATION, THERE WAS VISIBLE CORROSION IN THE BATTERY COMPARTMENT. ON TESTING, THE VIBRATOR MOTOR WAS NOT FUNCTIONAL. A REWIND, LOAD AND PRIME SEQUENCE WAS PERFORMED WITH NO LOSS OF PRIME. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO LOSS OF PRIME. A FORCE SENSOR CALIBRATION TEST REVEALED THAT THE FORCE SENSOR WAS OUT OF SPECIFICATION. A LEAK TEST WAS PERFORMED AND REVEALED A LEAK AT THE DISPLAY LENS. THE PUMP COVER WAS REMOVED FOR INVESTIGATION AND REVEALED MOISTURE INSIDE THE PUMP. CORROSION WAS NOTED ON THE INTERNAL BATTERY, THE FORCE SENSOR, FORCE SENSOR PLATE, FORCE SENSOR INTERNAL MOTOR GEARS AND THE VIBRATOR MOTOR.
RECALL # 2531779-03/24/2010-003-R.
THE PUMP HAS BEEN RETURNED AND WAS EVALUATED ON (B)(6) 2012 WITH THE FOLLOWING FINDINGS: THERE WAS CORROSION ON THE SPRING OF THE BATTERY CAP, INTERNAL BATTERY, FORCE SENSOR AND VIBRATOR MOTOR. REVIEW OF THE BLACK BOX REVEALED SEVERAL LOSS OF PRIME RECORDED. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |