CHARGING SYSTEM
Report
- Report Number
- 1627487-2012-02520
- Event Type
- Injury
- Date Received
- October 18, 2012
- Date of Event
- September 24, 2012
- Report Date
- September 24, 2012
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVALUATION: RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2012-02519. IT WAS REPORTED THE PT EXPERIENCES A HEATING SENSATION AT HER IPG SITE WHILE CHARGING. THE SJM REPRESENTATIVE ADVISED THE PT OF CHARGING PRECAUTIONS AND A NEW CHARGING SYSTEM WAS SENT TO THE PT. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CHARGING SYSTEM | SCS CHARGING SYSTEM | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3721 | 3509886 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Required Intervention | SCS LEAD: MODEL: 3189 (2)| IMPLANT DATE: |