FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2800730 · Received October 22, 2012

Report

Report Number
2032227-2012-07496
Event Type
Malfunction
Date Received
October 22, 2012
Date of Event
August 13, 2012
Report Date
October 17, 2012
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED AN ERROR DURING THE BASIC OCCLUSION TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED AN ERROR DURING THE MANUAL PRIME. THE CALLER STATED THAT ALARM OCCURRED WITH MULTIPLE INFUSION SETS. INSTRUCTED THE CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC MINIMED MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 36 YR