FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2800695 · Received October 22, 2012

Report

Report Number
3004209178-2012-90267
Event Type
Malfunction
Date Received
October 22, 2012
Date of Event
August 20, 2012
Report Date
October 17, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD SCRATCHED DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS, AND MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED AN ERROR DURING THE MANUAL PRIME PROCESS. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 11.7MMOL/L. ADVISED THE CALLER REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.