FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2800608 · Received October 22, 2012

Report

Report Number
3004209178-2012-90237
Event Type
Injury
Date Received
October 22, 2012
Date of Event
October 8, 2012
Report Date
October 10, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE, AND HIS BLOOD GLUCOSE WAS TREATED WITH AN INSULIN DRIP. TROUBLESHOOTING WAS PERFORMED. THE TIME, DATE, AND BOLUS WIZARD WERE CORRECT. ASSISTED THE CALLER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. PERFORMED A HIGH PRESSURE TEST AND PASSED. INSTRUCTED THE CUSTOMER TO CHANGE THE ENTIRE INFUSION SET AND RESERVOIR. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization