FDA Adverse Event Death Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2800554 · Received October 22, 2012

Report

Report Number
2649622-2012-15605
Event Type
Death
Date Received
October 22, 2012
Date of Event
October 5, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THE PROXIMAL CONDUCTOR WAS CUT, THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

AN IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS RETURNED TO THE MANUFACTURER FROM AN UNKNOWN SOURCE WITH NO INFORMATION. FURTHER REVIEW OF THE MANUFACTURER'S DATABASE INDICATED THE PATIENT IS DECEASED AND DIED APPROXIMATELY THREE WEEKS AFTER DEVICE SYSTEM WAS IMPLANTED. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076

Patients

Seq Age Sex Outcome Treatment
1 83 YR Death