HT COMMAND
Report
- Report Number
- 2024168-2012-06649
- Event Type
- Malfunction
- Date Received
- October 22, 2012
- Date of Event
- September 26, 2012
- Report Date
- September 26, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- DQX
- PMA / PMN Number
- K103101
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. DIL CATH: MINI TREK; SHEATH: 5F TERUMO 55 CM. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION. THE MINI TREK IS BEING FILED UNDER A SEPARATE MANUFACTURING REPORT NUMBER.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE RESISTANCE WITH THE CATHETER WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL, DIMENSIONAL, AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY TORTUOUS, HEAVILY CALCIFIED, AND 100% STENOSED ANTERIOR TIBIA ARTERY. A 2.0 X 20 MM MINI TREK BALLOON CATHETER WAS ADVANCED OVER THE GUIDE WIRE TO THE LESION WITHOUT RESISTANCE; HOWEVER, AS THE BALLOON CATHETER REACHED THE LESION, IT COULD NOT MOVE FORWARD OR BACKWARD OVER THE GUIDE WIRE. THE DEVICES WERE REMOVED AS A SINGLE UNIT. A NEW COMMAND GUIDE WIRE AND MINI TREK WERE USED SUCCESSFULLY. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HT COMMAND | GUIDE WIRE | DQX | AV-TEMECULA-CT | 2082001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |