FDA Adverse Event
Malfunction
Summary report: N
BREEZE2
MDR report key: 2800189
·
Received October 22, 2012
Report
- Report Number
- 1826988-2012-00558
- Event Type
- Malfunction
- Date Received
- October 22, 2012
- Date of Event
- September 1, 2012
- Report Date
- September 28, 2012
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K062347
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF 159MG/DL FROM HER BREEZE2 AND A READING OF 69MG/DL FROM ANOTHER METER. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE CUSTOMER WAS ADVISED TO RETURN THE TEST STRIPS FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BREEZE2 | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 1467A | 1T6190AA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |