FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2800184
·
Received October 12, 2012
Report
- Report Number
- 3008642652-2012-02682
- Event Type
- Malfunction
- Date Received
- October 12, 2012
- Date of Event
- September 22, 2012
- Report Date
- October 8, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK THERAPY ELECTRODE MESSAGES) WAS CONFIRMED. UPON INVESTIGATION THE WHITE PULSE WIRE WAS OPEN IN THE TRUNK CABLE INSULATION, WHICH CAUSED THE CHECK THERAPY MESSAGES. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE ON THE TRUNK CABLE. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS RECEIVING CHECK THERAPY ELECTRODE MESSAGES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |