LIFEVEST WCD 3100 SYSTEM
MDR report key: 2800143
·
Received October 12, 2012
Report
- Report Number
- 3008642652-2012-02683
- Event Type
- Malfunction
- Date Received
- October 12, 2012
- Date of Event
- September 29, 2012
- Report Date
- October 8, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESPONSE BUTTONS NOT WORKING) WAS CONFIRMED. UPON INVESTIGATION THE METAL SWITCHES WERE MISSING FROM THE FRONT AND REAR RESPONSE BUTTONS. THE ROOT CAUSE FOR THE DAMAGES RESPONSE BUTTONS COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED MONITOR RESPONSE BUTTONS. THE PT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) FEMALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER RESPONSE BUTTONS WEREN'T ACTIVATING HER MONITOR. THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |