FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2800127 · Received October 12, 2012

Report

Report Number
3008642652-2012-02706
Event Type
Malfunction
Date Received
October 12, 2012
Date of Event
September 15, 2012
Report Date
October 9, 2012
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF MONITOR SN (B)(4)6 IS CURRENTLY UNDERWAY. THE REPORTED PROBLEM (SCREECHING SOUND, WILL NOT POWER UP) HAS BEEN CONFIRMED. UPON EVAL THE MONITOR WOULD NOT POWER ON. UPON INVESTIGATION, THERE WAS A SEGMENTATION FAULT WHILE PERFORMING THE FLASH MEMORY TEST. THE CAUSE OF THE PROBLEM WAS ISOLATED TO FLASH MEMORY COMPONENTS U102 AND U105 ON COMPUTER ANALOG C/A BOARD SN (B)(4). A ROOT CAUSE ANALYSIS IS CURRENTLY UNDERWAY. A SUPPLEMENTAL REPORT WILL BE SENT UPON COMPLETION OF EVAL. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR. THE PT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WAS MAKING A SCREECHING SOUND AND WOULDN'T POWER UP. THE PT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 63 YR