FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 2800122
·
Received October 22, 2012
Report
- Report Number
- 3015876-2012-00782
- Event Type
- Malfunction
- Date Received
- October 22, 2012
- Date of Event
- September 17, 2012
- Report Date
- September 25, 2012
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K102972
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE BIPHASIC PCB ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A FAILURE OF AN INTEGRATED CIRCUIT CHIP, DESIGNATOR U12.
Description of Event or Problem · 1
IT WAS ORIGINALLY REPORTED THAT THE CUSTOMER'S DEVICE HAD ITS SERVICE INDICATOR ILLUMINATED AND HAD LOGGED AN EVENT CODE. AFTER FURTHER EVALUATION, IT WAS OBSERVED THAT THE CUSTOMER'S DEVICE WOULD NOT HAVE BEEN ABLE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC |