FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2800109 · Received October 12, 2012

Report

Report Number
3008642652-2012-02669
Event Type
Malfunction
Date Received
October 12, 2012
Date of Event
July 11, 2012
Report Date
October 5, 2012
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESETTING) HAS BEEN CONFIRMED. AS RECEIVED, THE MONITOR WAS RESETTING. THE CAUSE OF THE MONITOR NOT POWERING ON WAS A FLASH MEMORY FAILURE (COMPONENTS U102 AND U105) ON THE C/A BOARD SN (B)(4). THE FLASH MEMORY HAD AN INTERMITTENT BGA CONNECTION, WHICH WAS DISCOVERED THROUGH THE USEE OF A TARGET MEMORY TEST. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE COMPONENTS. THE PATIENT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) DISTRIBUTOR RETURNED A MONITOR INDICATING THAT THE A PATIENT'S MONITOR WAS RESETTING. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA