LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2012-02655
- Event Type
- Malfunction
- Date Received
- October 12, 2012
- Date of Event
- July 19, 2012
- Report Date
- October 4, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DOES NOT POWER ON) HAS BEEN CONFIRMED. AS RECEIVED, THE MONITOR WOULD NOT POWER ON. THE CAUSE OF THE INABILITY TO POWER ON WAS A FLASH MEMORY FAILURE (COMPONENTS U102 AND U105) ON THE C/A BOARD SN (B)(4). THE FLASH MEMORY HAD AN INTERMITTENT BGA CONNECTION, WHICH WAS DISCOVERED THROUGH THE USE OF A TARGET MEMORY TEST. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE COMPONENTS. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
A (B)(6) DISTRIBUTOR CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) PATIENT'S MONITOR WOULD NOT POWER ON. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |