FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2800028 · Received October 17, 2012

Report

Report Number
1720753-2012-08385
Event Type
Malfunction
Date Received
October 17, 2012
Date of Event
October 5, 2012
Report Date
October 17, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE POWER SUPPLY WAS ADJUSTED. THE BOARDS AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM HAD A COMMUNICATION ERROR AND LOCKED UP. THERE WAS NO PT INJURY OR DEATH REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1