GORE EXCLUDER AAA ENDOPROSTHESIS
Report
- Report Number
- 2017233-2012-00698
- Event Type
- Death
- Date Received
- October 11, 2012
- Date of Event
- June 16, 2011
- Report Date
- September 16, 2012
- Manufacturer
- W.L. GORE & ASSOCIATES
- Product Code
- MIH
- PMA / PMN Number
- P020004
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GR
- Reporter Occupation
- PHYSICIAN
Narratives
RESULTS: ADDITIONAL INFORMATION REGARDING SPECIFIC LOT/SERIAL NUMBERS AND PATIENT INFORMATION WAS REQUESTED BUT NOT AVAILABLE. THEREFORE, NO ANALYSIS OR REVIEW OF THE MANUFACTURING RECORDS COULD BE PERFORMED. CONCLUSIONS: THE ARTICLE DOES NOT SPECIFY OR INDICATE; AND GORE WAS UNABLE TO DETERMINED WHICH EVENT, IF ANY, EVENT A GORE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO. THE INSTRUCTIONS FOR USE FOR THE GORE EXCLUDER AAA ENDOPROSTHESIS STATES: ADVERSE EVENTS THAT MAY OCCUR AND / OR REQUIRE INTERVENTION INCLUDE BUT ARE NOT LIMITED TO; ENDOLEAK, COMPONENT MIGRATION, ANEURYSM RUPTURE AND DEATH. THE ARTICLE IDENTIFIED THE "MEAN" PATIENT AGE AND SEX TO BE A (B)(6) MALE.
AN ARTICLE PUBLISHED BY THE JOURNAL OF VASCULAR SURGERY, VOLUME 56, NUMBER 2 REVIEWED THE OUTCOMES OF ENDOVASCULAR REPAIR WITH SELECTIVE INTERNAL ILIAC ARTERY COVERAGE WITHOUT COIL EMBOLIZATION. KONSTANTINOS O. PAPAZOGLOU, MD, PHD, A GEORGE S. SFYROERAS, MD, PHD, B. NEOFYTOS ZAMBAS, MD, PHD, A DONSTANTINOS KNOSTANTINIDIS, MD, PHD, A STAVROS K. KAKKOS, MD, PHD, C AND MARIA MITKA, MD, A THESSALONIKI, ATHENS, AND PATRAS, GREECE. IN REVIEW OF THE PUBLISHED LITERATURE THE FOLLOWING FINDINGS WERE NOTED: BETWEEN (B)(6) 1997 AND (B)(6) 2010, 137 PATIENTS UNDERWENT ENDOVASCULAR ANEURYSM REPAIR (EVAR) OF THEIR ABDOMINAL AORTIC ANEURYSM (AAA) WHICH INCLUDED INTENTIONAL COVERAGE OF THE INTERNAL ILIAC ARTERY (IIA). IN 112 PATIENTS OF THE PATIENTS THERE WAS NO EMBOLIZATION (NO EMBOLIZATION [NE] GROUP) OF THE IIA AND IN 25 PATIENTS THE IIA WAS COIL EMBOLIZED (COIL EMBOLIZATION [CE] GROUP). COVERAGE OF THE IIA WAS BOTH UNILATERAL AND BI-LATERAL. ANATOMIC INDICATIONS FOR COVERAGE OF THE IIA WITHOUT COIL EMBOLIZATION INCLUDED PRESENCE OF ADEQUATE SEALING IN THE DISTAL 5 MM OF THE CIA, SEALING RING AT THE ORIGIN OF THE CIA, OR IIA DIAMETER <5 MM. PREOPERATIVE MEAN AAA SIZE WAS 60 - 14 MM, AND MEAN CIA DIAMETER WAS 38 - 13 MM. POSTOPERATIVE COMPUTED TOMOGRAPHY (CT) SCANNING WAS PERFORMED AT 1, 6, AND 12 MONTHS, AND YEARLY THEREAFTER, (130 MEN; (B)(6)). THE PATIENTS WERE FOLLOWED UP FOR 33-30 MONTHS. THE ARTICLE SPECIFIES THE FOLLOWING STENT GRAFTS WERE IMPLANTED IN THE 137 PATIENTS INVOLVED IN THIS STUDY: EXCLUDER (N 57), TALENT (N 53), ENDURANT (N10), ANEURX (N7), LIFEPATH (N3), ENDOLOGIX (N3), ENDOFIT (N3), AND COOK ZENITH (N1). THE ARTICLE DOES NOT SPECIFY WHICH DEVICE(S) ARE RELATED TO ANY INCIDENT REPORTED. THE RESULTS OF THE STUDY DEMONSTRATE THAT ENDOVASCULAR TREATMENT OF AORTOILIAC ANEURYSMS WITHOUT COIL EMBOLIZATION OF THE IIA IS SAFE AND EFFECTIVE, IS ASSOCIATED WITH A LOW INCIDENCE OF POSTOPERATIVE COMPLICATIONS AND REINTERVENTIONS, AND HAS ACCEPTABLE MIDTERM RESULTS. THE ARTICLE REPORTS THAT DURING FOLLOW-UP FOUR PATIENTS EXPIRED AND THE CAUSE OF DEATH WAS ANEURYSM RELATED. TWO PATIENTS HAD PROXIMAL MIGRATION OF THE ENDOGRAFT WITH ENDOLEAK; AND PRESENTED WITH ANEURYSM RUPTURE AT 45 AND 49 MONTHS POSTOPERATIVELY. THE PATIENT EXPIRED. THE THIRD PATIENT HAD A TYPE I ENDOLAK AFTER THE OPERATION AND UNDERWENT COIL EMBOLIZATION 3 MONTHS LATER; HOWEVER, THE ENDOLEAK PERSISTED. HE PRESENTED WITH ANEURYSM RUPTURE 20 MONTHS AFTER THE FIRST OPERATION. THE PATIENT EXPIRED. THE FOURTH PATIENT WAS REPORTEDLY DEBILITATED IN THE POST OPERATIVE SETTING, REQUIRING SURGERY FOR GLUTEAL SKIN NECROSIS RESULTING IN SEPSIS, END STAGE RENAL DISEASE AND GANGRENE. TO PRESERVE IS IIA, A BRANCHED GRAFT PROCEDURE WAS REQUIRED; BUT THE PATIENT WAS UNFIT FOR A COMPLICATED LONG PROCEDURE DUE TO HIS SEVERE CO-MORBIDITIES. THE PATIENT EXPIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GORE EXCLUDER AAA ENDOPROSTHESIS | MIH / SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | W.L. GORE & ASSOCIATES |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Death |