SROM STM ST,36+6L NK,11X16X150
Report
- Report Number
- 1818910-2012-22651
- Event Type
- Malfunction
- Date Received
- October 16, 2012
- Date of Event
- August 8, 2011
- Report Date
- October 10, 2012
- Manufacturer
- DEPUY IRELAND
- Product Code
- LPH
- PMA / PMN Number
- K954935
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
EXAMINATION OF THE REPORTED DEVICE WAS NOT POSSIBLE AS IT WAS NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER REPORTS AGAINST THE PRODUCT AND LOT CODE COMBINATION SINCE ITS RELEASE TO DISTRIBUTION. THE INVESTIGATION CAN DRAW NO CONCLUSION WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
PATIENT WAS REVISED TO ADDRESS PAIN. **UPDATE** (B)(4) 2012 - REVISION OPERATIVE REPORT WAS RECEIVED. THE PATIENT WAS REVISED TO ADDRESS PSEUDOTUMOR, CYST, SOFT TISSUE MASS, AND CORROSION WAS NOTED AT THE NECK TRUNNION TAPER LEVEL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SROM STM ST,36+6L NK,11X16X150 | TOTAL HIP REPLACEMENT | LPH | DEPUY IRELAND | 2769481 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |