FDA Adverse Event Malfunction Summary report: N

SROM STM ST,36+6L NK,11X16X150

MDR report key: 2792724 · Received October 16, 2012

Report

Report Number
1818910-2012-22651
Event Type
Malfunction
Date Received
October 16, 2012
Date of Event
August 8, 2011
Report Date
October 10, 2012
Manufacturer
DEPUY IRELAND
Product Code
LPH
PMA / PMN Number
K954935
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE REPORTED DEVICE WAS NOT POSSIBLE AS IT WAS NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER REPORTS AGAINST THE PRODUCT AND LOT CODE COMBINATION SINCE ITS RELEASE TO DISTRIBUTION. THE INVESTIGATION CAN DRAW NO CONCLUSION WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS PAIN. **UPDATE** (B)(4) 2012 - REVISION OPERATIVE REPORT WAS RECEIVED. THE PATIENT WAS REVISED TO ADDRESS PSEUDOTUMOR, CYST, SOFT TISSUE MASS, AND CORROSION WAS NOTED AT THE NECK TRUNNION TAPER LEVEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SROM STM ST,36+6L NK,11X16X150 TOTAL HIP REPLACEMENT LPH DEPUY IRELAND 2769481

Patients

Seq Age Sex Outcome Treatment
1 66 YR