FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
MDR report key: 2792676
·
Received October 16, 2012
Report
- Report Number
- 3015876-2012-00769
- Event Type
- Malfunction
- Date Received
- October 16, 2012
- Date of Event
- September 19, 2012
- Report Date
- September 19, 2012
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K103567
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL ISOLATED THE CAUSE FOR THE MALFUNCTION TO THE THERAPY PCB ASSEMBLY. PHYSIO REPLACED THE THERAPY PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S DEVICE WOULD NOT FUNCTION PROPERLY IN AED MODE. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | 15 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |