FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR

MDR report key: 2792676 · Received October 16, 2012

Report

Report Number
3015876-2012-00769
Event Type
Malfunction
Date Received
October 16, 2012
Date of Event
September 19, 2012
Report Date
September 19, 2012
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K103567
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL ISOLATED THE CAUSE FOR THE MALFUNCTION TO THE THERAPY PCB ASSEMBLY. PHYSIO REPLACED THE THERAPY PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S DEVICE WOULD NOT FUNCTION PROPERLY IN AED MODE. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.